  • Design of a Differential Electromagnetic Transducer for Use in IME System

    Byung-Seop SONG  Min-Kyu KIM  Young-Ho YOON  Sang-Heun LEE  Jin-Ho CHO  

     
    PAPER-Speech and Hearing

      Vol:
    E87-D No:5
      Page(s):
    1231-1237

    A differential electromagnetic transducer (DET) was implemented using micro electro mechanical system (MEMS) technology for use in an implantable middle ear (IME) system. The DET is designed to have good vibration efficiency and structure that can't be interfered by the external environmental magnetic field. In order to preserve the uniform vibration performance, the MEMS technology was introduced to manufacture the elastic membrane using polyimide that is softer than silicon. Using the finite element analysis (FEA), vibration characteristics are simulated and designed so that the resonance frequency of the membrane is closed to that of the middle ear. The results of the vibration experiments of the developed DET showed excellent results. We implemented the IME system using a DET and implanted it into a dog. This showed the IME system performed well in a living body.

  • A Transcutaneous Recharging System with the Function of Bi-directional Signal Transmission for Fully-Implantable Middle Ear Hearing Devices

    Il-Yong PARK  Hyung-Gyu LIM  Young-Ho YOON  Min-Kyu KIM  Byung-Seop SONG  Jin-Ho CHO  

     
    LETTER

      Vol:
    E89-A No:6
      Page(s):
    1692-1694

    In this paper, for the fully-implantable middle ear hearing devices (F-IMEHD), a transcutaneous recharging system that has the function of the bi-directional signal transmission with the implant module in a body as well as recharging battery has been designed and implemented. The electromagnetic coupling method using two coils has been adopted for the transfer of electrical power to recharge internal battery of the implant module. To increase the efficiency of power transfer, the switching frequency of recharging system is determined by the consideration of the resonance of LC tank circuits. The bidirectional signal transmission between the recharging system and the implant module has been designed through the on-off keying modulation of switching signal in the recharging system and the impedance variation of LC tank circuit in the implant module. Through the demonstration of the implemented system, it has been verified that the proposed system has the performance of bidirectional signal transmission with the implant module of F-IMEHDs as well as the battery recharging.

  • Telemetry Capsule for Pressure Monitoring in the Gastrointestinal Tract

    Ki-Won YOON  Sang-Hyo WOO  Jyung-Hyun LEE  Young-Ho YOON  Min-Kyu KIM  Chul-Ho WON  Hyun-Chul CHOI  Jin-Ho CHO  

     
    LETTER

      Vol:
    E89-A No:6
      Page(s):
    1699-1700

    In this paper, the pressure monitoring telemetry system has been designed and implemented for an accurate pressure measure-ment inside the gastrointestinal tract with minimizing pain and inconvenience. The system is composed of a miniaturized pres-sure measurement capsule and an external receiver. The per-formance of the telemetry capsule for monitoring pressure in the gastrointestinal tract is demonstrated by the results of animal in-vivo experiments.

  • New Method of Moving Control for Wireless Endoscopic Capsule Using Electrical Stimuli

    Hee-Joon PARK  Jyung-Hyun LEE  Yeon-Kwan MOON  Young-Ho YOON  Chul-Ho WON  Hyun-Chul CHOI  Jin-Ho CHO  

     
    PAPER

      Vol:
    E88-A No:6
      Page(s):
    1476-1480

    In order to control the moving speed of an endoscopic capsule in the human intestine, electrical stimulation method is proposed in this paper. The miniaturized endoscopic capsule with the function of various electrical stimulations has been designed and implemented. An in-vivo animal experiment has been performed to show the ability of controlling the movement speed of the endoscopic capsule according to the level of electrical stimulation. In-vivo experiments were performed by inserting the implemented capsule into a pig's intestinal tract. From the experimental results, the activation of peristaltic movement and the relationship between the moving speed of capsule and the stimulation amplitude could be found. It is shown that the moving speed of capsule in the intestine can be controlled by adjustment of the stimulation level applied in the capsule electrodes. The results of the in-vivo experiment verify that the degree of contraction in the intestinal tract is closely related with the level of stimulating electrical voltage, suggesting that the moving speed of capsule in the human gastrointestinal tract can be controlled by externally adjusting the amplitude of stimulating pulse signal.

  • Design of a Transcutaneous Infrared Remote Control for the Totally Implantable Middle Ear System

    Young-Ho YOON  Eui-Sung JUNG  Byung-Seop SONG  Sang-Heun LEE  Jin-Ho CHO  

     
    LETTER-Organic Molecular Electronics

      Vol:
    E88-C No:9
      Page(s):
    1896-1899

    An infrared (IR) transcutaneous remote control was designed for use in the totally implantable middle ear system. Considering the IR reflection, absorption and scattering effect of the skin, the required IR radiant intensity is calculated. After we have implemented the designed control, the transcutaneous operation experiment was carried out using a porcine skin.

  • Design of a Signal Processing Module with Various Filters Characteristics for Fully Implantable Middle Ear Hearing Devices

    Young-Ho YOON  Hyung-Gyu LIM  Jyung-Hyun LEE  Hee-Joon PARK  Il-Yong PARK  Min-Kyu KIM  Chul-Ho WON  Byung-Seop SONG  Jin-Ho CHO  

     
    LETTER

      Vol:
    E89-A No:6
      Page(s):
    1695-1698

    In this paper, the voice signal processing module has been designed using the micro processor for the use of fully implantable middle ear devices (F-IMEHD). The voice signal processing module for F-IMEHD should be designed to compensate for the hearing loss of hearing impaired person and have the flexibility for compensating various hearing threshold level. So, the voice signal processing module has been designed and implemented to present the various frequency characteristics using the low-power micro processor, MSP430F169. The different voice signal path to the inner ear entrance was considered so that two voice signal would be combined in-phase using an all pass filter with a constant time-delay to improve the vibration of the ossicles.