1-6hit |
Byung-Seop SONG Min-Kyu KIM Young-Ho YOON Sang-Heun LEE Jin-Ho CHO
A differential electromagnetic transducer (DET) was implemented using micro electro mechanical system (MEMS) technology for use in an implantable middle ear (IME) system. The DET is designed to have good vibration efficiency and structure that can't be interfered by the external environmental magnetic field. In order to preserve the uniform vibration performance, the MEMS technology was introduced to manufacture the elastic membrane using polyimide that is softer than silicon. Using the finite element analysis (FEA), vibration characteristics are simulated and designed so that the resonance frequency of the membrane is closed to that of the middle ear. The results of the vibration experiments of the developed DET showed excellent results. We implemented the IME system using a DET and implanted it into a dog. This showed the IME system performed well in a living body.
Il-Yong PARK Hyung-Gyu LIM Young-Ho YOON Min-Kyu KIM Byung-Seop SONG Jin-Ho CHO
In this paper, for the fully-implantable middle ear hearing devices (F-IMEHD), a transcutaneous recharging system that has the function of the bi-directional signal transmission with the implant module in a body as well as recharging battery has been designed and implemented. The electromagnetic coupling method using two coils has been adopted for the transfer of electrical power to recharge internal battery of the implant module. To increase the efficiency of power transfer, the switching frequency of recharging system is determined by the consideration of the resonance of LC tank circuits. The bidirectional signal transmission between the recharging system and the implant module has been designed through the on-off keying modulation of switching signal in the recharging system and the impedance variation of LC tank circuit in the implant module. Through the demonstration of the implemented system, it has been verified that the proposed system has the performance of bidirectional signal transmission with the implant module of F-IMEHDs as well as the battery recharging.
Ki-Won YOON Sang-Hyo WOO Jyung-Hyun LEE Young-Ho YOON Min-Kyu KIM Chul-Ho WON Hyun-Chul CHOI Jin-Ho CHO
In this paper, the pressure monitoring telemetry system has been designed and implemented for an accurate pressure measure-ment inside the gastrointestinal tract with minimizing pain and inconvenience. The system is composed of a miniaturized pres-sure measurement capsule and an external receiver. The per-formance of the telemetry capsule for monitoring pressure in the gastrointestinal tract is demonstrated by the results of animal in-vivo experiments.
Hee-Joon PARK Jyung-Hyun LEE Yeon-Kwan MOON Young-Ho YOON Chul-Ho WON Hyun-Chul CHOI Jin-Ho CHO
In order to control the moving speed of an endoscopic capsule in the human intestine, electrical stimulation method is proposed in this paper. The miniaturized endoscopic capsule with the function of various electrical stimulations has been designed and implemented. An in-vivo animal experiment has been performed to show the ability of controlling the movement speed of the endoscopic capsule according to the level of electrical stimulation. In-vivo experiments were performed by inserting the implemented capsule into a pig's intestinal tract. From the experimental results, the activation of peristaltic movement and the relationship between the moving speed of capsule and the stimulation amplitude could be found. It is shown that the moving speed of capsule in the intestine can be controlled by adjustment of the stimulation level applied in the capsule electrodes. The results of the in-vivo experiment verify that the degree of contraction in the intestinal tract is closely related with the level of stimulating electrical voltage, suggesting that the moving speed of capsule in the human gastrointestinal tract can be controlled by externally adjusting the amplitude of stimulating pulse signal.
Young-Ho YOON Eui-Sung JUNG Byung-Seop SONG Sang-Heun LEE Jin-Ho CHO
An infrared (IR) transcutaneous remote control was designed for use in the totally implantable middle ear system. Considering the IR reflection, absorption and scattering effect of the skin, the required IR radiant intensity is calculated. After we have implemented the designed control, the transcutaneous operation experiment was carried out using a porcine skin.
Young-Ho YOON Hyung-Gyu LIM Jyung-Hyun LEE Hee-Joon PARK Il-Yong PARK Min-Kyu KIM Chul-Ho WON Byung-Seop SONG Jin-Ho CHO
In this paper, the voice signal processing module has been designed using the micro processor for the use of fully implantable middle ear devices (F-IMEHD). The voice signal processing module for F-IMEHD should be designed to compensate for the hearing loss of hearing impaired person and have the flexibility for compensating various hearing threshold level. So, the voice signal processing module has been designed and implemented to present the various frequency characteristics using the low-power micro processor, MSP430F169. The different voice signal path to the inner ear entrance was considered so that two voice signal would be combined in-phase using an all pass filter with a constant time-delay to improve the vibration of the ossicles.